Inquiry (采购产品): Request for Medical Filter Information We are currently designing a vacuum system for an automated diagnostics instrument that will require a filter to remove possible biological contamination from air. The following is preliminary. System information: · Gas filtration (air) · Possible bleach vapor · 15 to 30 °F · 5 to 12 inches Hg vacuum · 10 to 20 liters/minute flow rate Filter requirements: · 0.22 micron absolute filtration (antimicrobial) · hydrophobic · maximum pressure drop 0.6 inches Hg at 90 liters/minute Selection criteria: 1. Cost effectiveness 2. Service/replacement life 3. Size Please provide information regarding possible filter solutions your company may be able to provide to meet the needs of our system. Thank you, David Combs **** Hidden Message ***** David Combs 10210 Genetic Center Dr. San Diego, CA 92121 USA Company: Gen-Probe Title: Sr. Mechanical Engineer Phone: 858 410 8342 Fax: 858 410 8302 E-mail: davidco@gen-probe.com **** Hidden Message ***** davidco@gen-probe.com