Inquiry (采购产品): 653 CS We need some technical information concerning the calibration/functional test of the 653 CS in the IN/SEC mode. Our Air Force calibration procedure has us set the meter calibrator to 207.5 mV RMS @ 10HZ and the 653 CS units to 200 (in/sec). The Air Force procedure states the outcome should be 180.0 nominal indication. The reading we get is 18.0. We believe the meter calibrator should be set to 2074.6 mV RMS to get a reading of 180.0 but have no way of proving this is correct. Can you help us resolve this issue? Thank You. Rick Rice **** Hidden Message ***** richard rice 09459 USA Company: Raytheon Technical Services E-mail: richard.rice@lakenheath.af.mil **** Hidden Message ***** richard.rice@lakenheath.af.mil